OPERATING INSTRUCTIONS 
 
 
18
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 SWEEP 
OPERATION 
 
1. Select LINEAR sweep by leaving the SWEEP LIN/LOG switch 
(23) in the released position or select LOG sweep by engaging the 
SWEEP LIN/LOG switch. 
2. Rotate COARSE FREQUENCY Control to minimum (CCW). 
3. Engage the RUN/SET switch to adjust the start and stop sweep 
frequencies. 
4. With the START/STOP switch set to START (released position), 
adjust the starting frequency of the sweep by adjusting the START 
control and observing the counter. 
5. Engage the START/STOP switch to the STOP position and adjust 
the stop sweep frequency by adjusting the STOP control. 
6. Release the RUN/SET switch for sweep operation. The sweep rate 
can be adjusted with the SWEEP TIME control. Clockwise 
rotation increases the sweep rate. 
 
If the output of the circuit to be tested is connected to the vertical 
scope input, and the GCV output to the horizontal, setting the scope to 
X-Y mode produces the amplitude vs. frequency plot. However, note 
that switching to LOG sweep still produces a linear display on the scope. 
This is because the horizontal sweeping signal, the internal log ramp, 
also becomes logarithmic when the sweep does. To view a true 
logarithmic graph, put the scope back in time base operation and use the 
Sweep output solely as a scope trigger. Use the scope's linear time base 
as a horizontal deflection source. 
 
USE OF FREQUENCY COUNTER WITH EXTERNAL 
SIGNALS 
 
 
The frequency counter display in the Model 4040 is 
normally used to monitor the internally generated output. 
However, it can also be used as a stand-alone frequency 
counter up to 30 MHz. 
 
1. Set the CNTR INT/EXT switch to EXT. 
2. Apply the signal to be measured to the EXT CNTR 
INPUT jack on the front panel. 
3. Obtain the desired display resolution by selecting the 
appropriate gate period using the GATE pushbutton 
switch. The resolution obtained with each of the four gate 
settings is as follows, given an input signal of exactly 2 
KHz. 
 
Gate Setting 
Reading 
Indicator 
0.01S 
   
 
  
 
 
 
 
 
 
2.0
KHz
0.1S
2.00
KHz
1S
2000
Hz
10S
2000.0
Hz