[ROUTe:]
The 
ROUTe
 subsystem controls switching and scanning operations for the
Microwave Switch modules in a switchbox. 
Subsystem Syntax
[ROUTe:]
CLOSe <
channel_list
>
CLOSe? <
channel_list
>
OPEN <
channel_list
>
OPEN? <
channel_list
>
SCAN <
channel_list
>
:MODE NONE | VOLT
:MODE?
CLOSe
[ROUTe:]CLOSe <
channel_list
>
  closes the Microwave Switch channels
specified in the channel_list.  The channel_list is in the form (@ccnn)
where cc = card number (00-99) and nn = channel number (00-04). 
Parameters
Parameter Name
Parameter Type
Range of Values
Default Value
channel_list
numeric
cc00 - cc04
N/A
Comments
Closing Channels: To close:
a single channel, use 
[ROUTe:]CLOSe (@ccnn)
;
multiple channels, use 
[ROUTe:]CLOSe (@ccnn,ccnn)
sequential channels, use 
[ROUTe:]CLOSe (@ccnn:ccnn)
;
a group of sequential channels, use 
[ROUTe:]CLOSe (@ccnn:ccnn,ccnn:ccnn)
;
or any combination.
Closure order for multiple channels with a single command is not
guaranteed.
Related Commands: 
[ROUTe:]OPEN
[ROUTe:]CLOSe?
 *RST Condition: All channels are open.
Example
Closing Microwave Switch Channels
This example closes channels 100 and 202 of a two-module switchbox (card
numbers 01-02).
CLOS   (@100,202)
100 closes channel 00 of card #1;
202 closes channel 02 of card #2.
58   HP E1368A/69A/70A Command Reference
Chapter 4