HP E1368A Manual De Usuario

Descargar
Página de 86
:CTYPe?
SYSTem:CTYPe? <
number
>
  returns the module (card) type of a selected
module in a switchbox.
Parameters
Parameter Name
Parameter Type
Range of Values
Default Value
number
numeric
01 - 99
N/A
Comments
Microwave Switch Module Model Number: 
The 
SYSTem:CTYPe? <
number
>
 command returns (for both modules):
HEWLETT-PACKARD,E1368A,0,A.01.00
Note: The 0 after the module number is the serial number of the module
(always 0).  A.01.00 is an example of the module revision code number.
Example
Reading the Model Number of a Card #1 Module
SYST:CTYP? 1
Returns the model number.
:ERRor?
SYSTem:ERRor?
  returns the error numbers and corresponding error
messages in the error queue of a switchbox.  See Appendix C for a listing of
the switchbox error numbers and messages.
Comments
Error Numbers/Messages in the Error Queue: Errors generated
by a switchbox store an error number and corresponding error
message in the error queue.  Each error message can be up to 255
characters long.
Clearing the Error Queue: An error number/message is removed
from the queue each time the 
SYSTem:ERRor?
 command is sent.
The errors are cleared first-in, first-out.  When the queue is empty,
each following 
SYSTem:ERRor?
 command returns 0, "No error".
To clear all error numbers/messages in the queue, execute the 
*CLS
command.
Maximum Error Numbers/Messages in the Error Queue: The
queue holds a maximum of 30 error numbers/messages for each
switchbox.  If the queue overflows, the last error number/message in
the queue is replaced by -350, "Too many errors".  The oldest error
numbers/messages remain in the queue and the most recent are lost.
*RST Condition: 
*RST
 does not clear the error queue. 
Example
Reading the Error Queue
SYST:ERR?
Query the error queue, read and
print the numbers/message.
66   HP E1368A/69A/70A Command Reference
Chapter 4