Hitachi Microscope & Magnifier S-4800 Manual De Usuario

Descargar
Página de 294
2 - 39
2.3.6.8  TILT COMPENSATION block    
 
Specimen tilting causes defocusing and magnification change to the tilt direction.   
Tilt compensation is the tool to correct the two phenomena. 
 
 
 
 
 
 
 
 
 
(1) Dynamic focus 
Dynamic Focus scans the focal length linked with the scanning positions.    When a specimen 
is tilted, the field of view that can be focused is small (particularly at low magnifications and at 
a short WD).    Dynamic Focus function allows you to focus the beam for the entire field of 
view. 
Refer to <3.9.9 Raster Rotation, Dynamic focus and Tilt Compensation> 
 
(2) Tilt compensation 
When a specimen is tilted, magnification along the tilting direction is [cos (tilting angle)] lower 
than that of a non-tilting direction.    As a result, the image appears to be contracted in the 
tilting direction. The image can be corrected at the magnification in all directions by using Tilt 
Compensation function. 
Refer to <3.9.9 Raster Rotation, Dynamic focus and Tilt Compensation> 
 
 
2.3.6.9  COLOR MIXING block    
 
The Color Mixing mode displays a color composite image.    Each of three colors (Red, Green or 
Blue) is assigned for two images signals for A (Left) and B (Right) screens, and displayed as a 
color composite image. 
This function is available only in Dual Screen mode. The color composite image is displayed on 
the right (B) screen. 
 
 
 
To display a color composite image, check the COLOR MIXING box and assign A or B signal (or 
OFF) for the three colors by selecting the appropriate colors and.    For example if the A signal is 
SE and the B signal is BSE, and A is assigned to Blue and B is assigned to both Red and Green, 
Dynamic focus check box 
Dynamic focus set / indication box 
Dynamic focus adjust slider 
Dynamic focus adjust button 
Tilt-magnification compensation check box 
Tilt-magnification compensation set / indication box 
Tilt-magnification compensation adjust slider 
Tilt-magnification compensation adjust button