3 - 9
(2) Type II (5-axis motorized stage): 
Click the EXC button on the control panel. The stage is then moved to the specimen exchange 
position, and the color of the indicator beside the button turns green.   
 
                                                       
 
Specimen exchange positions: X: 35.0mm 
Y: 55.0mm 
     
 
 
R: 0゜   
T: 0゜   
Z: 8.0mm 
 
NOTICE:  Do not repeat clicking the EXC button. The Stop button may become ineffective by it. 
 
NOTICE:  The specimen exchange position is not same as the stage home position. You need 
to move to the home position by clicking the HOME button after setting specimen. 
 
 
 
3.2.5    How to Set Specimen 
 
Before setting a specimen on the specimen stage, make sure that the height of the specimen is 
aligned with the height gauge mark. 
 
 
3.2.5.1 Extracting specimen from stage   
 
When a specimen is set inside the chamber, extract the specimen as follows.   
 
(1) Move stage to its specimen exchange position. 
After the color indicator becomes green, press the OPEN button on the specimen exchange 
chamber. 
                               
 
Once the exchange chamber will be evacuated and then, the gate valve will open. 
CAUTION
 
 
Start specimen exchange operation after the color of the indicator is changed green.