3.5.7.6    X, Y, R axes operation 
 
(1)   Manual operation using trackball 
Stage moves to the direction of rotating trackball. 
 
 
 
 
 
 
 
 
 
Amount of stage motion by ball rotation is linked with magnification.    To move long distance, 
it is advisable to lower the magnification.   
 
When you rotate ball pressing down one of upper buttons, stage moves only to Y direction. 
And pressing down one og lower buttons, only to X direction. 
 
Stage moving direction for ball rotation direction can be reversed. Check the Trackball / 
Joystick check box in the Reverse mode area of the Stage tab in the Optional setup dialog 
window. (Refer to <3.9.15 Optional Setting>) 
 
(2)    Entering Absolute or Relative Coordinates Value 
For X and Y axes, enter absolute coordinates for X and Y axes to XY Display / Input box and 
then, press ENTER key.   
For R axis, both absolute and relative value can be used. Select Abs (absolute) or Rel 
(relative) with Abs / Rel select buttons, and enter coordinate value to R Display / Input box 
and then, click Go button. Range of R coordinate is 0 to 360 for absolute and –180 to +180 for 
relative value. 
 
When the Rotation eucentric check box is checked, also X and Y axes moves so as the 
present field of view is kept after after the stage is rotated. 
 
   
 
(3) Continuous rotation (R axis) 
The R axis is driven with the Rotation.  Press 
 or 
  button to drive the R axis 
continuously or click the buttons to drive stepwise. 
A driving speed can be selected using the Speed slider in four steps. 
 
(4) Mouse operation on the scanning image 
All 5 axes operation by using mouse is available.   
Refer to <3.5.7.8 Stage mouse operation on the scanning image> 
 
Upper button 
Ball 
Lower button