SAFETY - 10
 
  WARNING 
 
Beware of 
electric shock
 
•  Voltages up to 100 V AC and 30 kV DC are used inside this 
instrument.    Do not touch inside or you may receive an electric shock. 
•  Do not remove the covers of column unit, control unit or power unit 
and do not touch internal parts or circuits while the instrument power 
is turned on.    There is a hazard of fatal or serious injury due to 
electric shock. 
•  Avoid removing the covers of the instrument.    If they must be 
removed, make sure to first turn off all power to the instrument. 
 
Electron Gun / Ion pump Baking 
•  The electron gun and ion pump are heated to about 250°C in order to 
keep the electron gun in a high vacuum.    Touching these sections 
would result in burns. 
 
Beware of high 
temperature
 
 
Rotary Pump 
•  The temperature of the oil rotary pump rises up to about 70°C, while it 
is operating.    Touching the pump may also result in burns. 
 
Beware of low 
temperature
 
 
Liquid Nitrogen 
•  When you replenish liquid nitrogen in the cold-trap of DP and the   
anti-contamination trap, be sure to wear eye-protection glasses and 
protective gloves to keep liquid nitrogen from splashing on the eyes or 
the skin.   
 
•  Anyone working on the electrical part of the system should follow the 
correct lock-out tag-out procedures for that location. 
•  The user should prepare a mains switch box on which the lock-out 
and the tag-out procedures are provided. 
•  Electrical hazards can occur if lock-out tag-out procedures are not 
followed. 
 
Magnetic Fields 
•  The ion-pump produces magnetic fields.    There are fringing magnetic 
fields of about 100 Gauss from the ion-pump and 10 Gauss at the 
outside of the ion-pump cover. 
 
Any person who has a pacemaker embedded is not allowed to 
approach the ion-pump. 
 
Do not bring near to the ion-pump wristwatches and magnetic cards 
that are susceptible to magnetic fields.