3.6.2    Preparing Images for Recording 
 
The source of image to be recorded is frozen image in the image memory except for direct 
photographing. 
 
(1)   Freezing an image 
 
(a) Using Run/Freeze button 
Scanning image will be frozen. When scanning speed is slow, scanning continues to end 
of the frame and then, frozen.   
Image size is 1280×960 pixels in Full screen. It is 640x480 pixels in small and dual 
screen mode. 
(b) Capturing an image 
An image is captured and frozen. Image size depends on the selected capture resolution 
(640x480, 1280x960, 2560x1920 or 5120x3840 pixels).   
(c) Direct photographing 
After photographing, the image is frozen in the image memory. Image size depends on 
the Capture Resolution setting in the Image Setup dialog window (640x480, 1280x960 or 
2560x1920 pixels). 
 
(2)   The source of image to be recorded 
Several commands or buttons are placed on plural area, the image sources for the commands 
are as follows. 
 
(a) When you use menu commands, buttons on the control panel, the scanning image    will be 
used as the image source. Image size depends the way of freezing, as mentioned above. 
(b) When you use buttons on the Captured Image window, the selected image on the window 
is used as the image source. Image size depends on the capture resolution setting. 
 
(3)   Embedding texts and graphics written on the image 
 
(a) Auto Data Display 
When the Embed Into Image box in the Data Display area on the Setup dialog window 
Record tab has been checked, the auto data display is embedded into the image data 
when the image is saved. If it is not checked, only the image data is saved. This Setting 
is also applied for Print, Copy and PCI Transfer commands. 
If the For Photograph box is checked, the position of the auto data display in the saved 
image is shifted slightly upward to ensure that it is properly framed in photographs. It is 
recommended not to heck the box if the saved image data will be used on the computer 
only.