3.8    At the End of Daily Use 
 
3.8.1    Turning High Voltage Off 
 
(1) Close opening dialog windows. Save data if necessary 
(2) Click the HV OFF button 
  on the control panel. 
 
 
3.8.2    Setting the Stage at the Specimen Exchange Position 
 
Set the all 5 axes of the specimen stage at the exchange position. 
(1) Type1 manual stage 
Release the stage lock.   
Set X = 25.0, Y = 25.0, Z = 8.0 Tilt = 0 and R = 0. 
(2) Type2 motorized stage 
Release the stage lock. 
Click EXC button on the control panel. Wait until the indicator of the button changed to green 
color. 
 
3.8.3.    Taking Out a Specimen 
 
(1) Press the OPEN button on the specimen exchange chamber. Once the exchange chamber 
will be evacuated and then, the gate valve will open. 
 
 
 
(2)   Turn the knob of the specimen exchange rod clockwise so as the Unlock mark comes 
upwards.  
                                                           
 
 
Insert the rod straightly and plug the two spring pins at the end of the rod to holes of the 
specimen holder.