3.11.4    Alignment of the Objective Lens Aperture 
 
(1) In the Beam Align mode, set the objective lens aperture at 2 or 3
Aperture diameter is 
  
 
opening 
1:100 
µ
  
 
opening 
2 and 3:50 
µ
m  
  
 
opening 
4:30 
µ
m  
 
(2) Adjust the brightness and contrast if the circular image becomes dark. 
 
(3) Adjust the X and Y knobs of the objective lens aperture so that the circular image is aligned to 
the center of the screen. 
 
 
 
 
 
 
 
 
3.11.5    Mechanical Alignment of the First Condenser Lens 
 
(1) In the Beam Alignment mode, set the Cond Lens 1 to 5 in the 
Column tab.   
 
(2) The circular image may become smaller and brighter. Adjust the 
brightness and contrast if necessary. 
 
(3) Adjust the alignment knobs of the first condenser lens so that 
the circular image is aligned to the center of the screen. 
 
 
3.11.6    Mechanical Alignment of the Second Condenser Lens 
 
(1) Click the Off button in the Alignment dialog window to terminate Beam Align mode.
   
 
(2) Check the Cond Lens 2 box (ON) in the Column tab.   
 
(3) Observe an image at a magnification of about 2000×, and 
adjust the brightness, contrast and focus. 
 
(4)  Place a point of interest in the center of the image display 
using the specimen stage.