1.1.3  Line Power Requirement 
 
(1)  Single-phase AC, 4 kVA, 50/60 Hz.   
When the supplied power line voltage is 100V, the instrument works by direct connection. 
When the supplied power line voltage is other than 100V (200V or others), use a step down 
transformer to step the power line voltage down to 100V. The transformer which meets to 
the standard of the country shall be prepared locally.   
 
Refer to 1.3.1 Power and Ground Wirings for connection method.   
 
NOTICE:     
(a) Line power fluctuation should be slow, and no abrupt fluctuation is allowable.   
(b) Continuous energization is required 
(c) Allowable line voltage fluctuation is 
±
10 % max. 
(2)  AC Power Distribution Unit should be located within 10 m of the power distribution board on 
the wall (since input AC cord is only 10 m long).   
(3)  Be sure to use an exclusive power switch on the distribution board which is provided by the 
user and feed the power through the switch.   
 
1.1.4  Grounding 
 
Connect the instrument to ground based on the standard of your country. 
Avoid sharing the ground terminal with other power equipments. Be sure to ground the 
instrument independently.  
 
1.1.5    Water Supply and Drain 
 
The following water supply conditions should be satisfied. 
Water flow rate 
:  1 to 1.5 L/min   
Water pressure 
:  50 to 100 kPa   
Water temperature  :  10 to 20 
°
C (temperature variation:    0.5 
°
C or less/10 min)   
It is recommended to supply water via a cooling water circulation unit.    When using city water, 
natural drainage is required.    If an appreciable amount of deposit is found, a proper water filter 
should be attached.    The drain port should be located at the floor level and water pressure must 
be free from pulsation in a short cycle.   
The water temperature for cooling the objective lens must be maintained so it is not more than 7 
°
C lower than the room temperature.    Condensation will form when the difference between the 
temperatures exceeds 7 
°
C.    And if the SEM is used for a long time under the condition where 
condensation forms, the objective lens area will become rusted and eventually be damaged, in 
which case the complete set of objective lens parts will require to be replaced. 
 
It is recommended to provide a water-leakage protection. Refer to Figure S-12 in Safety 
Summary section.