Hitachi Microscope & Magnifier S-4800 Manual De Usuario

Descargar
Página de 294
2 - 4
Fig. 2-4 shows the Type 1 specimen stage. 
 
 
 
 
Fig. 2-4    Specimen Stage (type I)   
 
 
[X] control  :  Moves specimen in X direction within a range from 0 to 50 mm. 
[Y] control  :  Moves specimen in Y direction within a range from 0 to 50 mm. 
[Z] control  :  Moves specimen in Z direction or shifts WD (Working Distance)   
within a range from 1.5 to 40 mm. 
(X-ray analysis position is at 15 mm working distance) 
[R] control  :  Rotates specimen through 360° (continuously variable) 
[T] control  :  Tilts specimen within a range from -5° to 70°. 
[Absorbed current measuring terminal]  :  Measures beam absorbed current on specimen. 
(Normally used as specimen applied voltage 
terminal) 
[Opening selector] 
:  Selects any of 4 aperture openings. 
[Fine opening control X knob] 
:  Effects fine control in the direction for introducing or 
taking out aperture. 
[Fine opening control Y knob] 
:  Effects fine control in the direction orthogonal to the 
above direction. 
[Stage lock switch] 
:  Used for locking the specimen stage to reduce 
vibration influence. 
Fine opening control X knob 
 
Opening selector 
 
Fine opening control Y knob 
 
[T] control 
 
 
Stage lock switch 
 
 
 
   
 
 
 
[X] control 
[Z] control 
 
 
 
 
[R] control 
 
Absorbed 
current 
measuring 
terminal 
 
 
[Y] control