Hitachi Microscope & Magnifier S-4800 Manual De Usuario

Descargar
Página de 294
2 - 27
When HV has been applied, click SET button after changing HV 
value.
   
• Set Ie to
 
  button shows a pull down list of Ie (emission current) value. 
Select a Ie value. 
• Vext indicator   
Extracting voltage value is indicated.   
• Flashing button 
The button further opens Flasing controller. This button is disabled 
while HV is applied.   
 
 
(3) Flashing controller 
 
 
• Intensity 
Selects flashing intensity. Use “2” for usual operation. 
• Execute button   
Executes flashing. Emission current caused by flashing will be 
indicated on the HV indicator about 2 seconds. Just after HV OFF 
or after flashing, 30 seconds wait time is necessary. If click Execute 
before 30 seconds passed, flashing will be executed after waiting 
rest of time. 
 
(4) Beam Monitor button 
 
 
Beam Monitor function is provided to reduce the tip noise, which is a low frequency noise 
caused by fluctuations of the emission current.    Dividing the image signal by a reference 
signal that is proportional to probe current can stabilize it.    It is recommended to keep Beam 
Monitor ON for normal operations. 
When the indicator above Beam Monitor button is green, the beam monitor function is working.   
If the indicator is OFF or blinking in red, click the button to turn it back on. 
 
Beam Monitor adjustment (adjustment of the reference voltage) is activated automatically at 
the start of ABCC function if Adjust Beam Monitor is checked in the ABCC area of the Image 
tab of the Setup dialog window.    Normally, check this box.    Uncheck this box in special 
cases such as when the Beam Monitor reference signal needs to trace the drift of the 
emission current; for example, during quantitative X-ray analysis using a probe current drift 
cancellation function, etc. 
 
To disable beam monitor function, uncheck Beam monitor Active in the SCAN MODE block of 
the Operation panel.