2.3.6.1  SIGNAL SELECT block 
 
 
   
 
 
 
 
 
(1) Signal select box 
Selects image signal for Full and Small screen mode and the A image of Dual screen mode. 
 
(2) Right half image Signal select box 
Selects image signal for the B image of Dual screen mode and the right half image of Split / 
Dual mag mode.   
 
(3) Signal addition indicator 
Controls adding two signals selected above Signal select boxes on the Dual screen mode. 
When you click the left side indicator, it becomes green color and the mixed of the two signals 
is displayed on the left (A) screen. If click the right side indicator, the mixed signal is displayed 
on the right (B) screen.    The signal addition is effective only on the Dual screen mode. 
 
(4) SE detector selection 
S-4800 SEM has two detectors, Upper and Lower. These have different characters and 
therefore result different image information. Also mixed signal of two detectors is available. 
Select with Mix / Upper / Lower buttons.   
Refer to <3.5.1 SE detector> 
 
(5) Signal Control tools 
Ratio of SE and BSE included in the upper detector output signal is controllable with the tools. 
When SE is selected, observing image will be of pure SE signal. 
When +BSE is selected, amount of BSE signal is controlled by BSE ratio selection box.     
Select a ratio from the pull down list clicking the arrow button. Low angle BSE will be detected 
with L.A0 to L.A100. With larger number, amount of SE is suppressed and results BSE richer 
image. H.A results high angle BSE image. 
Refer to <3.5.1.2 Signal Control>     
(6) GW BSE control (option) 
Opens control dialog window for GW type BSE detector. 
 
 
Signal select box 
Right half image Signal select box 
Signal addition indicator 
SE detector selection 
Signal control tools 
GW BSE control button 
BSE ratio selection box