2.3.9.1  Optics Tab 
 
The Optics tab is for setting operating condition of the electron optical column. 
 
(1)  ELECTRON BEAM Window 
Sets electron optical system conditions. 
 
 
 
Fig. 2.3-40    Setup Window – Optics Tab 
 
(a)  Vacc   
Sets and accelerating voltage.    This function is equivalent to the ELECTRON BEAM 
window on the Operation Panel. 
 
(b)  Emission Current 
Indicates the present emission current. 
 
(c)  Filament 
Changes the filament current and adjusts the temperature to which the filament is to be 
heated. The filament current should be adjusted to a point slightly below the saturation 
point for the emission current. 
The Auto Filament Saturation (AFS) function automatically sets an appropriate filament 
current.  AFS can be run in three modes, low, medium, and high, and can be used for 
the following purposes: