4.1.4 
4 - 15 
(7) 
Replace the aperture with a new one. 
 
(8) 
After the aperture has been replaced, perform Steps (4) - (5) in reverse to re-assemble the 
unit.    When turning the knob clockwise and inserting the BSED detector, turn it until the 
BSED detector catches the orifice unit and stops.    Notice that if the knob is turned too 
hard, the detector will disengage from the orifice unit and overshoot. 
 
(9) 
Press either the EVAC switch on the Evacuation Control Panel or the EVAC control button 
located on the right side of the Observation Screen Control Panel, and wait until the 
evacuation process finishes. 
 
 
 
 
 
 
EVAC Panel 
Control Panel 
 
Fig. 4.1-31    Evacuation Control Panel 
 
 
4.1.4    Maintenance for the Condenser Aperture 
 
The condenser aperture should be replaced once a year. 
Directions on how to replace a condenser aperture, prepared in 3D CG, are also available. 
For details, see 2.3.4, Menus. 
 
(1) 
If the accelerating voltage is ON, click OFF to shut it off. 
 
 
 
 
Fig. 4.1-32    Accelerating Voltage Setup Dialog 
 
(2) 
Either press the AIR switch on the Evacuation Control Panel or press the AIR evacuation 
control button located on the right side of the observation screen Control Panel.    Wait until 
the inside of the column reaches a complete AIR state.