Hanna Instruments Turbidity & Chlorine Measurements HI 93114 Manual De Usuario

Descargar
Página de 19
17
16
The main LCD will show
four dashes and “d 00”
is displayed on the
lower part, with the sec-
ond “0” flashing:
Using the  ,   and 
keys change the lower
row of the display to
show “d 11”:
Press the 
 key. This key is also used as a
toggle between USER (USR) and FACTORY
(FCT) programmed functions. Select the USER
mode (if necessary by pressing the 
 key
more than once):
Press ALT and CAL until four dashes are
displayed on the upper display and “#  fn” are
shown in the lower part of the LCD.
Using the   and   keys select the number
from 0 to 1 where the appropriate calibration
data have been stored.
Add the appropriate reagent into the blank
sample cuvet. Shake and allow a few sec-
onds for color to develop.
For the Free or Total Chlorine add the con-
tent of its respective packet:
Replace the cap and shake the cuvet. For
best results wait 2½ mins. for Total Chlorine.
Insert the reacted sample into the cuvet holder
and ensure that the notch on the cap is posi-
tioned securely into the groove. Press READ.
The meter will show “SIP” for a few seconds
and then the concentration:
Measurements in user-customized mode
Note: The meter must be calibrated for this
purpose. Follow the 2 point procedure
on page 26 before proceeding.
Turn the meter on, and press both ALT and
CAL buttons.
The meter will show “SIP” for a few seconds
and then a zero indication:
+
1 x