Enter the ambient air temperature at the transducer location. For optimum 
results, allow enough time (100 minutes) to ensure that the sensor is at 
equilibrium with the surrounding ambient temperature. Press 
ACCEPT
.
5.
Select the Liquid Depth method and enter the new level.
6.
Take a physical measurement of the liquid depth (level) and enter 
the value. Press 
ACCEPT
 when finished.
4.1.3.2 Sensor Height
This method requires entering the distance between the face of the ultrasonic 
sensor and the zero flow point in the primary device. The zero flow point in a 
primary device is the level at which flow ceases. In a round pipe the zero flow 
point would typically be the invert or bottom of the pipe. In a V-notch weir the 
zero flow point occurs when the liquid behind the weir is level with the bottom 
of the ‘V’. (There would still be liquid behind the weir plate but it would not be 
contributing to flow). Sensor Height calibration is generally used when:
Access to the primary device is difficult (such as confined space entry in a 
manhole) or
There is no liquid flowing during installation of the flow meter
1.
From the Main Menu, select 
OPTIONS>ADVANCED 
OPTIONS>CALIBRATION>ULTRASONIC SENSOR.
2.
Select Calibrate U-Sonic using the 
UP
 and 
DOWN
 keys. Press 
SELECT
.
3.
Select Standard as the type of Ultrasonic Transducer using the
 CHANGE 
CHOICE
 key. Press 
ACCEPT 
to continue.
The speed of sound in air varies with the temperature of the air. The 
ultrasonic sensor is equipped with temperature compensation to help 
eliminate the effect of temperature variation under normal site conditions.
4.
Enter the ambient air temperature at the transducer location. For optimum 
results, allow enough time (100 minutes) to ensure that the sensor is at 
equilibrium with the surrounding ambient temperature. Press 
ACCEPT
5.
Select the Sensor Height method and enter the new level.
6.
Enter the distance from the face of the transducer to the zero flow point of 
the primary device.
7.
Press 
ACCEPT
 when finished.
4.1.3.3 Setting the Invisible Range
1.
From the Main Menu, select 
OPTIONS>ADVANCED 
OPTIONS>CALIBRATION>ULTRASONIC SENSOR.
2.
Select the Invisible Range option using the 
UP
 and 
DOWN 
keys. Press 
SELECT
 to continue.
3.
Enter the Distance to End of the Invisible Range.