Conductivity Probe
8990options.fm
Section 5
5.5 Conductivity Probe
5.5.1 Conductivity Probe Connection
5.5.2 Conductivity Probe Programming
1.
From the Main Menu, select 
OPTIONS>ADVANCED OPTIONS>DATALOG.
2.
Highlight Select Inputs using the 
UP
 and 
DOWN 
keys. Press 
SELECT
.
3.
Highlight Conductivity (COND.) using the 
UP
 and 
DOWN
. Press 
SELECT
.
4.
Press 
CHANGE CHOICE
 to cycle between Logged and Not Logged, then 
press 
ACCEPT
.
5.
Enter a logging interval using the numeric keypad, then press 
ACCEPT
Valid logging intervals are shown on the status bar along the bottom edge 
of the display.
6.
Press 
CHANGE CHOICE 
to select the appropriate units (ms, µs). Press 
ACCEPT
 to continue.
7.
Select another channel to configure or press 
RETURN
 to back up one 
step. Press 
MAIN MENU
 to return to the Main Menu display.
5.5.3 Conductivity Temperature Programming
1.
From the Main Menu, select 
OPTIONS>ADVANCED OPTIONS>DATALOG.
2.
Highlight Select Inputs using the 
UP
 and 
DOWN 
keys and press 
SELECT
.
3.
Highlight Conductivity Temperature (COND. TEMP.) using the 
UP
 and 
DOWN 
keys, then press 
SELECT
.
4.
Press 
CHANGE CHOICE
 to cycle between Logged and Not Logged, then 
press 
ACCEPT
.
5.
Enter a logging interval using the numeric keypad, then press 
ACCEPT
Valid logging intervals are shown on the status bar along the bottom edge 
of the display.
6.
Press 
CHANGE CHOICE 
to select temperature units (°C, °F).
Press 
ACCEPT
Table 11 Conductivity Probe Wiring (J20)
Pin
Signal Description
Wire Color
A
Probe
Black
B
Probe
Red
C
RTD
White
D
RTD
Green