Hach 900 MAX Manual De Usuario

Descargar
Página de 160
Page 79
8990cm.fm
Communication Setup
Section 6
Communication Setup
Communication setup details connections and programming features of the: 
6.1 RS232 Cable
6.1.1 RS232 Connection
The RS232 connector is a serial input/output port for communicating with a 
sampler from an external device such as a Data Transfer Unit (DTU-II) or 
direct serial connection to a personal computer with application software.
Cable Required
RS232 Sampler/Flow Meter to PC Cable Assembly (Cat. No. 1727), 10 ft long, 
6-pin connector on one end, 9-pin D connector on the other end (9-pin to 
25-pin D type adapter included).
6.1.2 RS232 Programming
Note: Some early generation IBM 
compatibles may have a serial port 
that is not capable of 
communicating reliable at 19,200 
baud. If errors develop at high baud 
rates, try lowering the baud rate one 
step at a time (on both the sampler 
and Streamlogg II™, InSight
®
, or 
Vision
®
 software) until error free 
communications is achieved.
1.
From the Main Menu, select 
OPTIONS > ADVANCED OPTIONS > 
COMMUNICATIONS SETUP > RS232 SETUP
2.
Select a baud rate for data communications by pressing the
 CHANGE 
CHOICE
 key to cycle through the possible choices; 1200, 2400, 4800, 
9600 or 19,200 baud. The higher the baud rate setting, the faster data will 
transfer. Set the baud rate to the highest setting allowed by the computer. 
The baud rate must correspond to the baud rate selected in the software. 
Press 
ACCEPT
.
Note: Long runs of RS232 cable, especially if they are run near large motors or 
fluorescent lights can cause communication errors and may require a slower 
baud rate.
RS232 Serial Port (
)
Optional Modem (
)
4–20 mA Option (
)
Alarm Relays (
Analog Inputs (
Table 13 RS232 Connection
Pin
Signal Description
Wire Color
A
not used
White
B
ground
Blue
C
DSR
Yellow
D
RCD
Black
E
DTR
Red
F
TXD
Green