Analog Inputs
8990cm.fm
Section 6
There are a total of three analog input channels available on the sampler. 
These inputs accept 4–20 mA dc or -4.0 to +4.0 V dc analog signals. They 
can be logged and graphed and can also be used to trigger alarms, cause 
setpoint samples, and control 4–20 mA outputs. 
6.5.2 Analog Inputs Programming
Analog input channels can accept a signal from an external device. This 
signal may range from -4.0 V dc (min) to +4.0 V dc (max) or from 4 to 20 mA 
dc depending on the input selected. In some cases, input signals from certain 
devices may also fall somewhere within those ranges. For that reason, each 
analog input channel must be mapped to the minimum and maximum signal 
limits of the external device.
To map an external device to an analog input channel:
1.
From the Main Menu, select 
OPTIONS>ADVANCED OPTIONS>DATALOG.
2.
Highlight
 
Select Inputs
 
using the 
UP
 and 
DOWN 
keys and press 
SELECT
.
Note: If logging is enabled on any 
channel, then that channel will have 
an arrow in front of the channel 
name to signify the channel 
is logged.
3.
Highlight the analog channel to log using the 
UP
 and 
DOWN 
keys, then 
press 
SELECT
.
4.
Press 
CHANGE CHOICE
 to cycle between Logged and Not Logged, then 
press 
ACCEPT
.
5.
Enter a Logging Interval using the numeric keypad. Press 
ACCEPT
 to 
continue.
6.
Select Unit of measurement (ppm, ppb, afd, cfs, cfm, cfd, cms, cmm, cmh, 
cmd, gps, gpm, gph, lps, lpm, lph, or mgd).
7.
Enter Low Point.
8.
Enter High Point.
9.
Select another channel to configure, or press 
RETURN
 to back up one 
step. Press 
MAIN MENU
 to return to the Main Menu display.