3.5.2
Request for a temperature parameter, for example
The parameter index is not 3xh, thus the characters "from/to channel" and "receipt No." = 1, 1, 0 are contained in the control and long set.
Example: Request for the maximum selectable set point (SP H ) of the R2600/01 with address = 33 = 21h (see 4.1)
Request from master (control set, see 2.2):
Equipment response (long set, see 2.3):
See 2.4.2.2 for the contents of the function field (FF)
According to 4.1 and 2.4.4, the two characters of the data block (52h, 03h) result in:
SP H = 0352h = 850
If it is an equipment with marking B1 and configuration "sensor type" = 0 ... 6 and "sensor unit" = even number, the unit is 
°C 
(the configuration could be requested with PI = 32h and 33h).
68h
06h
06h
68h
21h
89
07h
01h
01h
00h
B3h
16h
FF
PI
PS
68h
08h
08h
68h
21h
„FF“
07h
01h
01h
00h
52h
03h
„PS“
16h
PI
2 characters
Data block