gohastings.com Hastings Entertainment Oxygen Equipment HFM-I-401 Manual De Usuario

Descargar
Página de 35
 
- 2 - 
1.1.4. Additional  Functions 
These instruments contain a number of functions in addition to reporting flow which include:  
•  Settable alarms and warnings with semiconductor switch outputs   
•  A digitally reported status of alarms and warnings such as overflow/underflow  
•  A flow totalizer to track the amount of gas added to a system 
•  A digitizing channel for an auxiliary analog signal   
•  An internal curve fitting routine for “fine tuning” the base calibration  
•  An alternate calibration set of 8 different ranges/gases 
   
1.2.  Specifications 
 
 
 
 
 
 
Performance 
HFM-I-401 HFM-I-405 
Full Scale Flow Ranges 
(in N
2
 0-10 slm up to 0-350 slm 
  0-100 slm up to 0-2500 slm 
Accuracy
1
 
Standard:  
± 1% full scale  Standard: 
± 1% full scale  
  
Optional:  ± (0.5% reading + 0.2%FS)  
Optional:  ± (0.5% reading + 0.2%FS)  
Repeatability 
± 0.1% of F.S. 
± 0.1% of F.S. 
Operating Temperature 
 -20 to 70°C 
 -20 to 70°C 
Warm up time 
30 min for optimum accuracy 
30 min for optimum accuracy 
  
2 min for ± 2% of full scale 
2 min for ± 2% of full scale 
Settling Time/Reponse 
Time 
< 2.5 seconds (to within 
± 2% of full scale) 
< 2.5 seconds (to within 
± 2% of full scale) 
Temperature Coefficient 
of Zero 
<  
±0.05% of Full Scale /°C 
<  
±0.05% of Full Scale /°C 
Temperature Coefficient 
of Span 
<  
±0.16% of reading/°C  
<  
±0.16% of reading/°C  
Operating Pressure -
Maximium 
Standard:  
500 psig Standard: 
 
500 psig 
  
Optional:  1500 psig 
Optional:  1000 psig 
Pressure Coefficient of 
Span 
< 0.01%of reading /psi      (N
2
, 0-1000 psig) 
< 0.01%of reading /psi      (N
2
, 0-1000 psig) 
Pressure Drop(N
2
@14.7 
psia) 
< 1.1 psi at full scale flow 
< 5.1 psi at full scale flow 
Attitude Sensitivity of 
Zero 
< 2% of F.S.  
< 2% of F.S.  
  
  
  
Electrical 
  
  
Power Requirements 
18-38 VDC, 3.5 watts(Ethernet) 2.5 
watts(RS232/485) 
18-38 VDC, 3.5 watts(Ethernet) 2.5 
watts(RS232/485) 
Analog Output  
Standard:  
4 – 20 mA  Standard: 
 
4 – 20 mA  
  
Optional:   0-10 VDC, 0-20 mA, 0-5 VDC, 1-5 VDC  Optional:   0-10 VDC, 0-20 mA, 0-5 VDC, 1-5 VDC 
Digital Output  
Standard:  
RS 232  Standard: 
 
RS 232  
  
Optional:  RS 485  
Optional:  RS 485  
  
Optional:  Ethernet  
Optional:  Ethernet  
Analog Connector  
Std: 
Terminal Block – PG 9 Cable Gland Std: 
Terminal Block – PG 9 Cable Gland 
  
Optional:  12 pin Circular Connector 
Optional:  12 pin Circular Connector 
WARNING 
 
Do not operate this instrument in excess of the specifications 
listed below.  Failure to heed this warning can result in serious 
personal injury and/or damage to the equipment.