- Before using the 
 Blood Glucose Monitoring System to test your blood glucose, please read all the 
instructions and conduct all the tests including the quality control test (Refer to page 36.)
- Please do the quality control test regularly to make sure the test results are accurate.
- The 
 Blood Glucose Meter can only be used with the 
 Blood Glucose Test Strips. Other 
brand strips should not be used under any circumstances. The use of other brands of strips may give 
inaccurate results.
- The 
 Blood Glucose Monitoring System is intended for in vitro diagnostic use only. The testing 
result is calibrated to be plasma equivalent to testing with fresh capillary whole blood samples from the 
fingertip, palm and forearm.
- The 
 Blood Glucose Monitoring System is intended for self-testing and professional use. It should 
not be used to diagnose diabetes mellitus.
 Blood Glucose Monitoring System has not been validated for use on neonates. Therefore, it's not 
intended for use on neonates.
 Blood Glucose Monitoring System has not been used for arterial blood testing yet, so it is not 
suitable to be used for arterial blood testing.
- Follow all environmental protection regulations when disposing of batteries.
GE100
GE100
GE100
GE100
GE100
GE100
GE100
- Do not perform a blood glucose test when USB cable is connected to the Blood Glucose Meter.
If the GE100 Blood Glucose Meter and Test Strips are exposed to a substantial change in temperature, 
please wait 30 minutes before measurement.
Caution
 
5
4
Precaution
A minimum of 1.0µL of blood is needed for testing with the 
 Blood Glucose Monitoring System. 
Blood sample size above 3.0µL might contaminate the test strip port and the meter.
Blood sample size below 1.0µL might cause inaccurate test result or might not start the meter. In this 
case, repeat the test with a new test strip.
GE100
CAUTION
- The 
 Blood Glucose Meter will be set to mg/dL by default when sold in the United States. 
When your display shows " mmol/L " during setting or test, please contact customer service. 
Use of the wrong unit of measure may cause incorrect treatment.  
GE100
Sample Size Example
- Avoid contact with dripping or splashing liquids.
- The minimum blood sample size to test using the 
 Blood Glucose Monitoring System is 1.0µL. 
GE100
1.0µL         1.5µL              2.0µL             3.0µL            4.0µL