Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001
1
Plug in the 41 °C calibration probe.
2
Gain access to the adjustment potentiometers VR4 and VR5, by removing the control board from the humidifier
front panel.
3
With the Chamber Temp button held down turn on the power switch. Display should show -1-
4
Using the Chamber Temp button step through to number -4-
5
Starting with a vertical bar display adjust VR4 so the display changes to an all 
8 display. If VR4 is rotated too
far the display becomes all horizontal bars (||.||.||. thru 
8.8.8. to ===). The correct point is where all the
segments of the all 
8 display are of equal brightness.
6
Using the Chamber Temp button step to number -5-
7
Remove the 41 °C calibration probe.
8
Adjust VR5 for a display reading of zero. (0)
(This value may drift over time. A deviation of up to ± 2.0 will not affect circuit performance.)
9
Calibration is complete; do performance check (§ 6.0)