Fisher & Paykel MR730 Manual De Usuario

Descargar
Página de 64
54
Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001
C2
MR700 S
ERIES 
F
AULT 
F
LOW 
C
HART 
F
OR 
N
G
C
ONDITION
Check Voltage 
Rating Of Unit 
Open Unit and check if 
thermal cut out under 
heaterplate has tripped
Reset thermal cut out and find reason 
for fault 
Check for mains voltage 
at Connector J5 on power 
board
Check connectors on power 
board are in correct locations and 
secure, check fuses (F1 - F4), 
check voltage on connector J7of 
power PCB, mains switch on, 
connectors on rear of control 
board and continuity of 
transformer, connector J8-1& 3 
(Primary) and connector J9-1 & 2 
+ 3 & 4 (Secondaries).Check 
voltage on transformer 
secondary  J9.
Replace power board 
and test
Replace control board 
and test
Return unit for repair
Perform calibration and 
performance checks as 
outlined in Sections 3.0 and 
4.0 of the Technical Manual 
before re-instatement of unit
Check mains lead and 
plug
Repair faulty component 
and test
Assemble and test
NO
YES
YES
GOOD
NO
NO
YES
YES
BAD
NO
YES