Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001
1.0
MR700 S
ERIES 
G
ENERAL 
S
PECIFICATIONS
_________________________
This documentation defines the technical specifications of the MR730, MR720, MR700 and MR480 respiratory
humidifiers, it also includes maintenance and repair procedures.
The MR730/720 models are Heated Wire humidifiers with the option of disabling the Heater Wire.
The MR720 has a lower alarm limit and an extended warmup mode for use in Anaesthesia.
The MR700 model is exclusively a Heated Wire humidifier.
The MR480 model is a non Heated Wire humidifier.
1.1
E
LECTRICAL 
S
PECIFICATIONS
1.1.1
M
ODELS 
MR730/720/700
Supply Voltage
230 ± 25 V
127 
±
 12 V
115 V
100 ± 10 V
Supply Frequency
50 or 60 Hz
Supply Current
1.0 A maximum at 230 V
1.9 A maximum at 127 V
2.0 A maximum at 115 V
2.4 A maximum at 100 V
Heater Wire Capacity
60 W
1.1.2
M
ODEL 
MR480
Supply Voltage:
230 ± 25 V
115 V
100 ±10 V
Supply Frequency
50 or 60 Hz
Supply Current
0.7 A maximum at 230 V
1.4 A maximum at 115 V
1.6 A maximum at 100 V
1.1.3
C
OMMON 
E
LECTRICAL 
S
PECIFICATIONS 
(A
LL 
M
ODELS
)
Heaterplate Power
150 W
Heaterplate Overheat Protector
Operates at 118 ± 6 °C
1.2
S
TANDARDS AND 
A
PPROVALS
Classification under:
CSA-C22.2 No.125 , UL2601
IEC601-1, AS3200.1, EN60601-1
Class 1
Type B
Drip Proof
Continuous Operation
Not to be used in the presence of flammable anaesthetics.
1.3
T
EMPERATURE 
C
ONTROL
PID (Proportional Integral Derivative) control of Heater Wire (in Heated Wire mode operation only) and heaterplate.
1.4
T
EMPERATURE 
S
ENSING
Temperature measurement using standard characteristic thermistors (YSI 400 compatible) mounted in fast response
temperature probes, located at the patient end of the delivery tube and at the humidifying chamber outlet.