Campbell Manufacturing CR10 Manual De Usuario

Descargar
Página de 253
SECTION 7.  MEASUREMENT PROGRAMMING EXAMPLES
7-18
FIGURE 7.16-3.  Hook up to AVW1
Program:  AVW1 & CR10 USED TO
MEASURE 1 GEOKON VIBRATING WIRE
SENSOR.
*
1
Table 1 Programs 
01:
60
Sec. Execution Interval 
01:
P4
Excite,Delay,Volt(SE) 
01:
1
Rep 
02:
15
2500 mV fast Range 
03:
1
IN Chan 
04:
1
Excite all reps w/EXchan 1 
05:
1
Delay (units .01sec) 
06: 2500
mV Excitation 
07:
1
Loc [:TEMP     ] 
08:
.001
Mult 
09:
0
Offset 
02:
P55
Polynomial 
01:
1
Rep 
02:
1
X Loc TEMP
03:
1
F(X) Loc [:TEMP     ] 
04:
-104.78
C0 
05:
378.11
C1 
06:
-611.59
C2 
07:
544.27
C3 
08:
-240.91
C4 
09:
43.089
C5 
03:
P28
Vibrating Wire (SE) 
01:
1
Rep 
02:
2
IN Chan 
03:
1
Excite all reps w/EXchan 1 
04:
24
Starting Freq. (units=100 Hz) 
05:
32
End Freq. (units=100 Hz) 
06:
500
No. of Cycles 
07:
0
Rep delay (units=.01sec) 
08:
2
Loc [:PRESSURE ] 
09:
-34.836
Mult