DB1 Liquid Level Measurement Sensor 
 
 
 
Case 3 
 
 
 
 
PortSet(3,1) 
'Open valve to atmosphere 
   EndSelect 
 
   Delay(0,150,MSEC) 
 
 
 
Sum = 0 
 
 
 
For j = 1 To 25 
'Take 25 pressure readings for each tube and atmosphere 
 
 
 
 
BrFull(mV, 1, mV25,4, VX1, 1, 2500, true, true, 0, _60Hz, 1, 0) 
    Sum 
Sum 
mV 
 
 
 
Next j 
   Valve(i)= 
Sum/25 'Average the 25 readings from upper and lower tube and atmosphere 
   PortSet(1,0) 
'Close valve to lower tube 
   PortSet(2,0) 
'Close Valve to upper tube 
   PortSet(3,0) 
'Close Valve to atmosphere 
 
 
Next i 
 
 
 
Delta_mV = Valve(2) - Valve(1) 
  Mult 
Delta_H/Delta_mV 
 
 
Offset = (Valve(3) * Mult) 
 
 
WtrLvl = Valve(2)* Mult – Offset  'units of WtrLvl are same units as Delta_H (a constant) 
 
  CallTable 
Table1 
 NextScan 
 
EndProg 
 
5.2  CR10(X) Example 
Input Location Usage: 
LOCATION   LABEL 
#1:  
Liquid Level 
#4: 
Multiplier   
#5: Offset 
#6: mv/V 
#7: Sum 
#8: 
Avg. Level  
#9: mV/V#1 
#10:  
mV/V#2 
#11:  
mV/V#3 
#12:  
Delta V 
#13:  
Delta h 
* Table 1 Programs 
  01:  60 
  Sec. Execution Interval  
 
1:  Do (P86) 
  1:  2 
  Call Subroutine 2  
 
2:  Do (P86) 
  1:  42 
  Set high Port 2