Bacharach Snifit 50 Manual De Propietario

Descargar
Página de 2
- 2 -
Figure 1. Snifi t 50
- 3 -
Figure 2. Calibration Equipment Setup
- 4 -
If you’re not sure about the quality of the surrounding 
air, you can apply a blend of Oxygen/Nitrogen gas to 
the sensor as described under Calibration.
If rapid temperature changes are encountered (such 
as from freezing to room temperatures) during the 
operation of the analyzer, the operator must allow the 
analyzer to stabilize at the ambient temperature for at 
least 2 minutes before taking a measurement.
Checking for CO
Important!  Ensure that the sensor grille at the rear of 
the instrument is unobstructed and open to the atmo-
sphere. A quick instrument check can be performed by 
allowing the smoke of a blown-out match to enter into 
the sensor grille. This should cause the indicated CO 
level to increase.
After zeroing the Snifi t, simply hold the instrument in 
the area that you suspect the presence of CO gas. If CO 
is present, the instrument will indicate the concentrA-
tion of CO in ppm on its display.
For more extensive CO investigations and testing in 
areas of higher temperatures, the Monoxor
®
 II is the 
preferred instrument.
Backlight
Pressing the Backlight button will illuminate the 
display for a period of 8 minutes, or until it is manually 
turned off by again pressing the Backlight button.
Overrange
If the CO level exceeds 1999 ppm, the 
displayed reading is replaced by the 
number “1”.
Low Battery
When the Snifi t’s 9V battery is nearing 
the end of its useful life, 
LO BAT
 will 
Calibration
Important!  For accurate Snifi t operation, calibration 
every 6 months is recommended.
To calibrate the CO sensor, you will need the equip-
ment listed under Accessories. Note that the Zero Gas 
Cylinder is needed only if you’re unsure about the qual-
ity of the surrounding air for zeroing purpose.
Calibrate the sensor to a known concentration of CO 
gas as follows:
1.  Assemble the calibration equipment per Figure 2.
2.  Zero the instrument as previously described under 
Zeroing the Sensor
  If necessary, you can use the calibration cup to apply 
a blend of Oxygen/Nitrogen gas directly over the sen-
sor by attaching a zero gas cylinder to the regulator 
and adjusting the regulator knob for a fl ow rate of 2 
SCFH.
 MONOXIDE ANALYZER
 MONOXIDE ANALYZER
ppm CO
CARBON
CARBON
Snifit
model 50
model 50
Backlight
On
Off
Span Adjust
-Power-
ZERO
ADJUST
(Recessed)
SPAN
ADJUST
POT.
BACKLIGHT
ON/OFF
POWER ON
POWER OFF
DISPLAY
LO BAT
LOW
BATTERY
INDICATION
1
2
3
4
Parts Shown:
1. CO Span Gas Cylinder, 51-1994
2. Regulator, 03-4318*
3. Tubing, 03-6351*
4. Flowmeter, 06-6163*
5. Calibration Cup Assembly, 23-2156**
*  Part of Calibration Kit 24-7059
** Part of Tool / Cup Kit 19-3242
5
PUSH CALIBRATION CUP 
OVER SENSOR HOUSING
LO BAT
Specifi cations
Gas Monitored .............Carbon Monoxide (CO)
Range ...........................0  to 1999 ppm
Resolution ....................1  ppm
Accuracy ......................± (5% of reading + 5 ppm)
Response ......................
30 sec. to 90% of fi nal value
Power ...........................9V battery
Battery Life .................1500 hours minimum
Temperature Range ....32 to 104 °F (0 to 40 °C)
Case Material ..............High impact plastic
Weight ..........................0.25 lb (0.11 kg) without battery
Dimensions ..................5.5" L x 2.0" W x 1.6" H 
(140 x 51 x 41 mm)
Operation
Turning the Snifi t 
On & Off
The Snifi t is turned on 
and off by pressing the 
Power On
 and Power 
Off
 buttons, respec-
tively. If the instru-
ment is not manually 
turned off, it will au-
tomatically shut itself 
off after approximately 
35 minutes.
Zeroing the 
Sensor
Turn on the Snifi t and allow the displayed reading to 
stabilize before proceeding – approximately 30 seconds. 
Then with the instrument sampling fresh air (air that 
is free of CO), use a screwdriver to adjust the Zero 
Adjust
 potentiometer at the top of the instrument for a 
reading of 0 ±1 ppm.
appear in the upper left hand corner of the display. 
Although you can continue using the instrument under 
this condition, you should replace the battery as soon 
as possible to ensure accurate CO readings.