Bacharach Hi Flow Sampler Manual De Propietario
Hi Flow Sampler
Operation
Instruction 0055-9017
21
3.14.2.2
Manual 1-Stage Mode
In the Manual 1-Stage mode the user starts a measurement at a high
flow rate and then waits for a stable leak rate reading to occur. During
this time the flow rate can be changed using the Speed function on the
Main Screen. After a stable leak rate reading is obtained, the user can
this time the flow rate can be changed using the Speed function on the
Main Screen. After a stable leak rate reading is obtained, the user can
manually save all final measured and calculated values in memory using
the Save function on the Expanded Main Screen. This mode is used
when more control of the leak rate measurement process is required.
the Save function on the Expanded Main Screen. This mode is used
when more control of the leak rate measurement process is required.
Select Manual 1-Stage operation as follows:
1. From the Expanded Main Screen, select Menu > Operating Mode.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
C a l i b r a t i o n
S e t F l o w U n i t s
M e n u M o d e
O p e r a t i n g M o d e
A c c e s s R e c o r d s
[ 0 0 0 1 ] ( M )
S - -
2. Highlight Manual 1-Stage, and then press the I/O key.
0 2 / 2 1 / 0 3 0 9 : 4 5 : 3 0
A u t o m a t i c 2 - S t a g e
M a n u a l 2 - S t a g e
M a n u a l 1 - S t a g e
[ 0 0 0 5 ] ( M )
S - -
3. Press ESC to return to the Main Screen. Observe that (M) should
appear at the bottom of the screen