2 X 2 SCSI MATRIX SWITCH
p. Set Switch (Device) Number:
example command: <cr>//|ODxx<cr>
xx = ASCII representation of decimal value from 01 through 32
report: none
q. Set “Lock/Unlock Front Panel” Password:
example command: <cr>//|OPxxxx<cr>
xxxx: ASCII 4-digit decimal number from 0000 to 9999
report: none
r. Set Temperature Limit:
example command: <cr>//|OT<cr>
xx: ASCII 2-digit decimal number from 15 to 55 (degrees C)
report: none
s. Report Revision of SCSI Switch’s Firmware:
example command: <cr>//|R<cr>
example report: 2059722
t. Lock Serial Ports:
example command: <cr>//|OLxxxx<cr>
xxxx: password set with OP command (ASCII 4-digit decimal number 
from 0000 to 9999)
example report: “//|OL1 pass” if password entered correctly, “//|OL0 fail”
otherwise
u. Unlock Serial Ports:
example command: <cr>//|OUxxxx<cr>
xxxx: password set with OP command (ASCII 4-digit decimal number 
from 0000 to 9999)
example report: “//|OU1 pass” if password entered correctly, 
“//|OU0 fail” 
otherwise
v. Report Serial-Port Lock Status:
example command: <cr>//|OS<cr>
example report: “//|OS1 lock” if the serial ports are locked, 
“//|OS0 open” 
otherwise