CHAPTER
Technical Information
15
37
Glucometer
®
E
SPRIT
™ System
(Whole Blood Referenced)
1. Accuracy: A patient-use clinical study was
performed at diabetes care centers in three 
different locations. Glucose levels were measured
with the Glucometer E
SPRIT
System on over
300 fresh capillary blood specimens by 61 persons
with diabetes, and six healthcare professionals.
The results were compared with the whole blood
glucose results of the YSI 2300 Stat Plus (Yellow
Springs Instrument Co., Inc., Yellow Springs, OH), 
a glucose oxidase/electrode system. 
Figure 1 on the following page shows the
comparison of the lay diabetic Glucometer 
E
SPRIT
System results with the whole blood 
glucose results. The comparison of the healthcare
professional results with the whole blood glucose
results is shown in 
Figure 2.
36
CHAPTER
15
Technical Information
REPLACEMENT PRODUCTS:
When calling or writing for supplies, be sure to include the
number with the name of the replacement part or product.
Part Number
Item
40030032
3-volt Lithium Battery (DL or
CR 2016) — 2 (Available at most stores)
99939679
Glucometer
®
E
SPRIT
™ User Guide
50184147
Glucometer
®
E
SPRIT
™ Quick Reference Guide
✽ 
Part Numbers subject to change without notice.
Product
Number
Product Name
3616
Glucometer
®
E
SPRIT
™ Test Sensor Discs — 50s 
3649
Glucometer
®
E
SPRIT
™ Low Control
3650
Glucometer
®
E
SPRIT
™ Normal Control
3651
Glucometer
®
E
SPRIT
™ High Control
6541
Microlet™ Automatic Lancing Device
5970
Glucolanz™ Lancets (200s)
6555
Microlet™ Endcaps — Regular (10s)
6559
Microlet™ Endcaps — Super (10s)
Replacement products may be obtained from Diabetes
Care Centers, or your local pharmacy.