Selecting Lens Type
Use optional accessory close-up lens for small area measurement. See
instruction manual for close-up lens for the placement of one. If close-up lens is
to use, measurement value is required for calibration of lens transmission factor.
Since calibration value varies depending on lens type, lens type has to be set in
this instrument in advance. Erroneous setting causes incorrect measurement.
Below table shows lens to set and setting for this instrument.
instruction manual for close-up lens for the placement of one. If close-up lens is
to use, measurement value is required for calibration of lens transmission factor.
Since calibration value varies depending on lens type, lens type has to be set in
this instrument in advance. Erroneous setting causes incorrect measurement.
Below table shows lens to set and setting for this instrument.
Lens to Set
No lens
Close-Up lens
No.107
Close-Up lens
No.122
Arbitrary lens like optional accessory ND
filter (1/10) CS-A6, ND Filter (1/100) CS-A7,
Combination of close-up lens and ND filter
Setting
STANDARD
No.107
No.122
USER1
∗
USER2
∗
USER3
∗
∗ Setting at the shipment from factory : STANDARD
∗ Use standard accessory data management software CS-S10w Standard or optional
CS-S10w Professional to avail more lens type like USER1, USER2 and USER3, and
register calibration value for lens other than close-up lens No.107 and No.122 set in
USER1, USER2 or USER3. See instruction manual for data management software for
details.
register calibration value for lens other than close-up lens No.107 and No.122 set in
USER1, USER2 or USER3. See instruction manual for data management software for
details.
Operation Procedure
1.
Press
ESC
key when menu or target
value setting menu appears.
Measurement screen appears on LCD screen.
2.
Make sure that [ ] appears on left
bottom indicating this instrument
has not been set for SHIFT mode.
If [SFT] appears instead, it means that
instrument is to operate in SHIFT mode.
Press
instrument is to operate in SHIFT mode.
Press
SHIFT key to cancel SHIFT mode.
2
3
1, 7
6
4, 5
O
B
JE
C
T
IV
E
LE
N
S
CH00:
<
DEFAULT
>
Lv 20.80
cd/m
2
x 0.4476
y 0.4477
<
MODE
>
SINGLE
[ ] [ 2° ]
<
ANGL
>
1° VIEW
<
SPD
>
AUTO
<
SYNC
>
200.00Hz
<
LENS
>
STANDARD
<
MEM
>
AUTO NUM
<
MEMORY DATA
>
M
OOO
:(NO DATA)
Lv ------
x ------
y ------