Konica Minolta Chroma Meter CS-200 Manual De Usuario

Descargar
Página de 124
88
Me
asu
rem
ent
Measurement
1.
  Decide whether you use close-up 
lens (optional) or not according to 
measuring object size and distance. 
See the table below for details on measuring distance 
and measuring area. If you set close-up lens, lens 
type setting is required in this instrument. (see p.40)
2.
  Slide measuring angle selector 
according to size of object and select 
either aperture1˚, 0.2˚, or 0.1˚.
∗ The measuring angle select is covered with holding cap 
CS-A24 (see p.8) at shipment from factory.
Pressing with finger cushion, slide 
measuring angle selector slowly.
Do not operate measuring angle selector during 
measurement. If measuring angle is switched during 
measurement, you may fail to perform measurement 
or to obtain correct measurement value.
3.
  Rotate diopter adjustment ring on 
finder for diopter adjustment. 
Make sure that aperture (black circle indicating 
measurement area) looks clear when 
observing object through finder. (See p.17.)
4.
   Rotate focus adjustment ring for 
objective lens for that purpose. 
Make sure that object around aperture looks 
clear when observing object through finder. 
Only measuring area for measuring object must 
be placed in aperture. If extra area, which is 
not measuring object, is included in aperture, 
you cannot perform correct measurement.
Operation Procedure
0.2°
0.1°
(unit:mm)
Minimum measuring area Maximum measuring area Minimum measuring area Maximum measuring area Measuring area at 500 mm Measuring area at1000 mm
(Measuring angle) 1
˚ 0.2˚ 0.1˚
1
˚ 0.2˚ 0.1˚
1
˚ 0.2˚ 0.1˚
1
˚ 0.2˚ 0.1˚
1
˚ 0.2˚ 0.1˚
1
˚ 0.2˚ 0.1˚
Without a Close-Up Lens 4.7
1.0
0.5
296
8.5
ø1.7 ø0.9 ø17.7 ø3.6 ø1.8
Close-up Lens No.122
2.2
0.5
0.3
4.6
1.0
0.5
128
240
Close-up Lens No.107
0.8
0.2
0.1
1.1
0.3
0.2
43
52
∗Measuring distance is the distance from the front edge of the metal lens barrel or close-up lens ring.
Measuring distance and measuring area