8842A
Instruction Manual
4-6
f4-04.wmf
Figure 4-4. Wire Ohms Measurement
Table 4-1. Ohms Test Current
RANGE
TEST CURRENT
FULL SCALE VOLTAGE
20
1 mA
0.02V
200
1 mA
0.2V
2 k
1 mA
2.0V
20 k
100 
µ
A
2.0V
200 k
10 
µ
A
2.0V
2000 k
µ
A
10.0V
20 M
500 nA
10.0V