PASCO Specialty & Mfg. PASCO Scientific SODIUM ION SELECTIVE ELECTRODE CI-6734 Manual De Usuario

Descargar
Página de 22
Sodium Ion Selective Electrode
012–06615A
10
Electrode Response
Plotting the electrode potential against the sodium concentration on semilogarithmic paper results in a straight line
with a slope of about 59 mV per decade. (Refer to Figure 1.)
The time needed to reach 99% of the stable electrode potential reading, the electrode response time, varies from
one minute or less for sodium concentrations above 1.0X10
-5 
M to several minutes near the detection limit. (Refer
to Figure 4.)
Figure 4
Typical sodium electrode time response to step changes in NaCl
Limits of Detection
The upper limit of detection in pure sodium solutions is 1 M. In the presence of other ions, the upper limit of detection
is above l0
-1 
M sodium, but two factors influence this upper limit. Both the possibility of a liquid junction potential
developing at the reference electrode and the salt extraction effect influence this upper limit. Some salts may extract
into the electrode membrane at high salt concentrations, causing deviation from the theoretical response. Either
dilute samples between l M and l0
-1 
M or calibrate the electrode at 4 or 5 intermediate points.
Free sodium ion concentration down to l.0Xl0
-6 
M or 0.1 ppm can be measured in basic solutions. For measurements
below l0
-5 
M or 1 ppm, use plastic lab-ware (and low level procedures) since a significant pickup of sodium may
occur from glassware due to removal from container walls.
pH Effects
The electrode response to sodium ions is greatly influenced by the pH of the solution. Hydrogen ion interferes with
measurements of low level sodium ion measurements, although the electrode can be used over a wide pH range.
(See Figure 5.)
-75
-50
-25
0
+25
+50
+75
+100
+125
1
2
3
4
10
-3 
M to 10
-2 
M NaCl
10
-3 
M to 10
-4 
M NaCl
10
-3 
M to 10
-5 
M NaCl
10
-3 
M to 10
-6 
M NaCl
time (minutes)
electrode
potential
(mV)