Fluke 943-27 Manual De Usuario

Descargar
Página de 45
General Information
Application
1
 
1-3 
 
Table 1-2.  General Specifications for the 943-227-15 Digital Preamplifier 
Feature: Specification: 
Preamplifier Power 
+ 15 v @ 20 mA 
Preamplifier Input 
DC current, 1E-12 to 1E-3 amperes from 943-27 Current Mode Beta 
Scintillation detector 
Preamplifier Output 
943-227-15, RS232C; 
943-227-15VL, Modified RS232C serial data for transmission distances up 
to 5,280 feet.  
Electronic Accuracy 
± 10% of the input current 
Response Time 
Decade / Time: 
1 /  80 sec 
2 / 32 sec 
3 / 16 sec 
4 /  4 sec 
5 /  2 sec 
6 / 0.9 sec 
7 / 0.2 sec 
8 / 0.2 sec 
 
Dimensions  
(H x W x D) 
943-227-15 or 943-227-15VL: 
8.0 in. x10.0 in. x 4.0 in. 
(20.3 cm x 25.4 cm x 10.2 cm) 
 
943-227-15-M1 or 943-227-15VL-M1: 
8.0 in. x10.0 in. x 6.0 in. 
(20.3 cm x 25.4 cm x 15.2 cm) 
 
Weight 
Approximately 8 lbs. (3.6 kg) 
Housing 
Gasketed, NEMA 4, Steel enclosure with hinged cover 
Operating Temperature 
32 °F to 122 °F 
0 °C to + 50 °C 
Relative Humidity 
0 to 99 % non-condensing 
Compatible Detectors 
Victoreen 943-27 Current Mode Beta scintillation detector only 
Detector Interface 
Detector cables enter thru “Seal Grip” fittings; Mating SHV receptacle 
provided for HV; Mating BNC receptacle provided for Signal  
Cable Requirements 
Victoreen Loop Communications: 6 conductors, 18 AWG (2, TwPr, 2 Single 
conductor) with overall shield.  For max transmission distance, 16 AWG is 
recommended; 
RS232C communications:  8 Conductors, 18 AWG (4 TwPr) with overall 
shield. 
Modification 
Information 
For the Model 943-227-15VL-M1, refer to Modification description provided 
in Appendix E