Fisher & Paykel MR810 Manual De Usuario

Descargar
Página de 43
 
 
MR810 Technical Manual Revision C  REF 185042601
 
17 
 
5
5.2 S
EE 
M
ANUAL 
C
ODES
 
When the See Manual Indicator is illuminated the humidifier will switch off the heater plate and heater-
wire. Pressing the Temperature Setting button displays the error code on the Temperature Setting 
Indicators, as explained in Table 3. 
Table 3: Indicator codes and their meaning 
Temperature 
Setting 
Indicators 
See 
manual 
Indicator 
Fault Description 
Section 
 
 
Microprocessor failure, replace PCB. 
6.4 
 
 
Button fault.  Check buttons are correctly seated. 
6.7.1 
 
 
Heater plate primary thermistor fault (short or open 
circuit). Check the thermistor is working correctly, 
replace if necessary. 
6.5.4 
 
 
Heater plate secondary thermistor fault (short or 
open circuit). Check the thermistor is working 
correctly, replace if necessary. 
6.5.4 
 
 
Ambient thermistor fault (short or open circuit), 
check heater-wire adaptor, replace if necessary. 
6.6 
 
 
Chamber Sense thermistor fault (short or open 
circuit), check heater-wire adaptor, replace if 
necessary. 
6.6 
 
 
Heater-wire relay fault (short or open circuit).  PCB 
faulty, service or replace PCB as necessary. 
6.4 
 
 
Heater control triac error (Heater-wire triac short 
circuit or heater plate triac open/short circuit).  
Service triac circuitry, replace PCB if necessary. 
6.4 
  
Heater-wire 
Clip 
Circuit error.  Service the 
hardware clip circuit, replace PCB if necessary. 
6.4 
 
 
Failed factory production testing.  Device should be 
returned to your Fisher & Paykel Healthcare 
representative. 
 
 
Temperature Setting Indicator description 
  
Indicator  off 
  
Indicator  flashing 
  
Indicator  on