MR810 Technical Manual Revision C  REF 185042601
 
 
3
3 O
PERATION AND 
C
ONTROLS
 
3.1 A
PPLICATION
 
The MR810 Respiratory Humidifier is designed to heat and humidify respiratory gases delivered to the 
patient via a face, nasal, or oral mask.  The inspired gas is passed through the chamber of water where it is 
heated and humidified.   An optional heated breathing circuit (heated via a wire inside the circuit), can be 
used to minimise condensate in the breathing circuit and maximise humidity delivery to the patient; this is 
recommended. 
The MR810 humidifier consists of two heaters: (a) a heater plate which the humidification chamber slides 
onto, and (b) an optional heater-wire in the breathing circuit. 
3.2 W
ARNINGS AND 
C
AUTIONS
 
WARNING: 
Ensure probe port caps are inserted into probe ports of the breathing circuit. 
Caution is required when delivering gases that are not body temperature saturated. 
The use of breathing circuits, chambers or other accessories which are NOT 
approved by Fisher & Paykel Healthcare may impair performance or compromise 
safety. 
Refer to operating instructions for each accessory. 
Ensure maintenance of grounding integrity by connection to a hospital grade 
receptacle. 
Ensure that the humidifier is always positioned lower than the patient’s airway. 
Ensure the humidifier is securely mounted. 
The operation of high frequency surgical apparatus, short wave or microwave 
equipment in the vicinity of the humidifier may adversely affect its function.  If this 
occurs the humidifier should be removed from the vicinity of such devices. 
Visually inspect accessories for damage before use. 
Hot surfaces may exceed 75 °C. 
Do not use flammable anaesthetics. 
Always disconnect supply before servicing. 
Caution 
The MR810 does not deliver body temperature saturated gases.