88  Stratos LV/LV-T Technical Manual   
2.8 Mode 
Switching 
Mode switching prevents the conduction of paroxysmal 
atrial tachycardias to the ventricle.  Therefore, after sensing 
an atrial tachycardia while in activated mode switching, the 
CRT-P automatically switches to an atrial-controlled R-
mode.  Like the programmed atrial-controlled P-mode, the 
corresponding R-modes can be programmed: 
Table 21. Mode Switching 
Programmed P mode 
Programmed R mode in case of 
sensed atrial tachycardias 
DDD DDI 
DDD DDIR 
DDDR DDIR 
VDD VDI 
VDD VDIR 
VDDR VDIR 
DDTA DDI 
DDTA DDIR 
DDTAR DDIR 
 
The Mode Switching algorithm causes the CRT-P to change 
pacing modes when a programmed number of atrial intervals 
(X) out of 8 consecutive atrial intervals (p-p) are faster than the 
programmed mode switch intervention rate (X out of 8).  X is 
programmable from 3 to 8.  The rate at which an atrial interval is 
determined to signify an atrial tachyarrhythmia is called the 
mode switch intervention rate.  The mode switch intervention 
rate is programmable from 100…(10)…250 bpm.