dEC1 
Selects the decimal point position 
(Options: 0, 0.0, 0.00, 0.000). 
 
C     0.00 
InC 
Selects the increment size. 
(Options: 1, 2, 5 and 10). 
 
inC   10 
CAP 
Sets the scale capacity.  The maximum value 
for the scale is as specified for the scale.  For 
example 60kg.   
 
 
P2 CAL 
CAL 
Starts the calibration procedure.  See the 
details of calibration procedure in section 10. 
 
 
 
The sub-parameter “
P3 Pro” is used to view the calibration constant used by the 
scale, view the A/D counts and set the scale back to default settings.  This is 
useful if it is not possible to have a good calibration of a scale.  It will show the 
load cell is causing the A/D to increment correctly through the complete range of 
weights needed and you can recall the default settings for the scale.   The “
trl” 
value can be modified to cause the scale calibration to change to a known 
amount without the need for putting weights on the scale.  
 
 
 
trl 
Display will show a value that corresponds to 
the load cell calibration.  If the value is 
increased the scale calibration will be 
decreased a similar percentage. 
   
 
CoUnt 
Display will show the counts from the A/D 
converter.  The displayed count should 
increase as the weight is added to the load 
cell up to and exceeding the capacity of scale.   
Typical range for these counts are: 
No load  20,000 
Maximum   120,000 
 
 
P3 Pro 
rESEt 
Scale will reset to factory default settings.  
Press  [Zero/Enter] to continue or [Print/Esc] 
to abort. 
 
 
• 
Once the value is set, press [Print/Esc] to return to the sub-parameter.  
 
• 
Pressing [Print/Esc] again will return the scale to normal weighing. 
 