9.5 Remote Commands and Error Codes
The following tables list all commands, and their descriptions, that
are accepted by the calibrator. 
Table 5: Common Commands
Command Description
*CLS 
(Clear status.) Clears the error queue.  
*IDN?
Identification query. Returns the manufacturer, model
number, and firmware revision level of the Calibrator.  
*RST 
Resets the calibrator to the power up state.  
Table 6: Calibrator Commands
Command Description 
 
DAMP 
Turns Damp on or off.  
DAMP? 
Returns if DAMP is on/off  
DISPLAY 
Turns on/off the displays specified in the command  
DISPLAY? 
Returns which displays are on/off  
FAULT? 
Returns the most recent error code   
ERROR_LOOP 
Turns loop power on or off in percent error mode   
ERROR_LOOP? 
Returns the current state of loop power in error mode  
ERROR_MODE 
Turns percent error mode on or off   
ERROR_MODE? 
Returns whether percent error mode is on or off   
ERROR_PORT 
Set the pressure port for percent error mode  
ERROR_PORT? 
Returns the pressure port for percent error mode  
FUNC 
Sets the display mode as specified in the command  
FUNC? 
Returns the current mode of the upper, middle, and
lower display  
HI_ERR 
Sets the 100% of span limit for percent error mode   
HI_ERR? 
Returns the 100% of span limit for percent error mode  
LOCAL
Returns user to manual operation of the calibrator  
LOCKOUT 
Locks out the keypad of the calibrator in remote opera-
tion  
LO_ERR 
Sets the 0% of span limit for percent error mode   
LO_ERR 
Returns the 0% of span limit for percent error mode  
PRES_UNIT 
Set the pressure unit for the indicated display  
PRES_UNIT? 
Returns the pressure from the indicated display  
REMOTE 
Puts the calibrator in remote mode  
27