Page 41 
8.6 
ANSI test results 
FastFacts 8.6: ANSI test results 
 
In 1996 tests: Full-on average gain 
j
 is measured at 50 dB SPL, average gain at 
RTP 
k
 is usually (but not always) measured at 60 dB SPL and the frequency 
response 
l
 is run at 60 dB SPL for linear aids or at 50 dB SPL for AGC aids. 
 
In 2003 tests: Full-on average gain is measured with a 50 dB input SPL, average 
gain at RTS is always measured at 60 dB input SPL and the frequency response 
curve is always run at 60 dB input SPL.