16  Insertion Gain 
16.1  Section overview 
FastFacts 16.1: Insertion gain 
16.2  Insertion gain in SPL 
FastFacts 16.2: Insertion gain in SPL 
Insertion gain is the difference between aided and unaided ear canal SPL (REIG = 
REAG - REUG). It inherently assumes a sound-field audiogram. For non-linear 
hearing aids, REIG is stimulus-dependent and the dependency is specific to each 
hearing aid. Insertion gain fitting methods ignore these facts. Speechmap is 
recommended for fitting all non-linear hearing instruments. 
 
Refer to 
10: Real-Ear Measurement Setup
 and 
11: REM Screen Setup
 for 
terminology and information on preparing the system for use and for proper 
positioning of the client and the probe tube. 
This section covers the following topics: 
16.2: Insertion gain in SPL 
16.6: REAR measurement procedure 
16.3: Insertion gain in HL 
16.7:SII calculation in Insertion gain 
16.4: Audiometric data entry 
 
16.5: REUR measurement procedure 
See 
10: Real-Ear Measurement Setup
 and 
11: REM Screen Setup
 
1)  Press <Tests>, then highlight & <PICK> [Insertion gain] from the REM tests list. 
2)  Highlight and <PICK> [Instrument] to chose from BTE, ITE, ITC, CIC, Body. 
This changes the CORFIG used for 2cc targets. 
3)  Highlight and <PICK> fitting method to chose from FIG6, NAL-NL1, NAL-R, 
POGO II, Berger, Libby, User, None