RM500SL User’s Guide Version 2.8  
Page 73 
17.6  Using Speechmap 
FastFacts 17.6: Using Speechmap 
The Speechmap fitting environment provides a number of distinctly different 
stimuli. Speech and noise are analyzed in 1/3 octave bands and (except for pink 
noise) data are presented as 3 curves. The top curve is the level exceeded 1% of 
the time (speech peaks), the lower curve is the level exceeded 70% of the time 
(speech valleys) and the middle curve is the average. Starting a test causes the 
signal to be presented continuously - the speech signals run in a loop - and the 
curves are updated frequently to show any changes you make to the hearing 
instrument. Pressing <Continue> restarts the speech passage and the curves are 
then calculated for the entire passage, producing an accurate display of the 
speech region and LTASS, and an accurate SII calculation. When using live 
speech, a <Freeze curve> button captures short-term speech features. See 
17.10: 
SII calculation in Speechmap
 and 
19:  Speechmap Technical 
Details. 
Before starting, please review 
10: Real-Ear Measurement Setup
17.2:Speechmap 
facts
17.9: REM or S-REM mode
v  To access Speechmap, press <Tests>, then highlight & <PICK> 
[Speechmap] from the REM tests list. 
v  To change screen setup, see 
11: REM Screen Setup
, and 
17.9: REM or S-
REM mode
v  To set up hearing instrument type, highlight & <PICK> [Instrument], then 
highlight & <PICK> BTE, ITE, ITC, CIC, Open, Body, CROS or FM. This 
changes microphone location effects and the RECD used in S-REM and 2cc 
targets, and the reference microphone used for FM and CROS. 
v  To enter audiometric data, see 
18.2: Assessment data entry
v  To run a test, highlight and <PICK> one of the 4 REAR buttons. Then 
highlight and <PICK> [Stimulus] to select the stimulus type. Highlight and 
<PICK> [Level] to select stimulus level. Press <Continue> to run the test and 
display the data. See, 
17.7: Screen tour
 & 
18: Speechmap Fitting Procedures