18  Speechmap Fitting Procedures 
18.1  Section overview 
FastFacts 18.1: Speechmap fitting procedures 
18.2  Assessment data entry 
FastFacts 18.2: Assessment data entry 
Threshold data and parameters may be entered by scanning a barcode on a 
previous RM500SL or Verifit printout or manually using a mouse, the keypad or 
the numeric keypad on an external keyboard. See 
2.6: Mouse, keyboard, barcode 
scanner
, and 
3.2: Input device operation.
 
To enter data from a barcode see 
3.3: Barcode data input
 and: 
1)  Press the barcode trigger and center the red line on the barcode from a 
distance of 6 - 8”. The scanner will beep and the Barcode Entry poster opens. 
2)  Highlight and <PICK> [Done] to accept 
the data indicated on the poster. 
To enter assessment data manually: 
1)  Highlight and <PICK> [Audiometry]. 
2)  Highlight and <PICK> an item to toggle its 
setting or generate a list box. Age can be 
changed only if Targets = [DSL child]. See 
18.3: Assessment data choices - 1
 and 
18.4: Assessment data choices - 2
3)  Press <Continue> to proceed to data entry. 
4)  Enter HL threshold using a mouse,keypad or keyboard. Left click the mouse on 
the audiogram to enter a point. Click again to delete the point. On the keypad 
the arrow keys change frequency and level and <PICK> enters or deletes a 
threshold point. On a keyboard enter HL values on the numeric keypad. 
5)  Press <Continue> when all points have been entered. 
6)  Repeat the process for Bone conduction, UCL and RECD if [Enter] has been 
selected for any. If RECD is set to [Measure], see 
15: RECD (real-ear to 
coupler difference)
The procedures in this section assume a familiarity with 
10: Real-Ear 
Measurement Setup
11: REM Screen Setup
 and 
17: Speechmap ®
This section covers the following topics: 
18.2: Assessment data entry 
18.7: Fitting to targets for mid-level speech 
18.3: Assessment data choices 
18.8: Open fittings in Speechmap 
18.4:Assessment data choices - 2 
18.9: FM fitting and verification 
18.6:Fitting to targets for loud sounds   
Related sections are 
17: Speechmap ®
 and 
19: Speechmap Technical Details