Page 82 
18.7  Fitting to targets for mid-level speech 
FastFacts 18.7: Fitting to targets for mid-level speech 
18.8  Open fittings in Speechmap 
FastFacts 18.8: Open fittings in Speechmap 
6)  Select REAR 3 and set stimulus to [Speech-std(1)] or [Speech-std(2)] and level 
to 70. Elongated 
+
 symbols show the target range for 70 dB speech LTASS. 
7)  Adjust gain for mid-level sounds so that the middle curve (the LTASS) falls 
within the target range. Press <Continue> to run the complete passage. Repeat 
as necessary. Achieving this goal may require adjusting gain for soft speech, 
which will reduce its audibility. This may be a necessary compromise with some 
hearing instruments and for clients with a very narrow dynamic range. 
 
REM mode should be used for open and vented fittings. There are no special real-
ear SPL targets for open fittings - it doesn’t matter how the sound reaches the TM, 
the perception is the same. However, the sound that escapes from a vent or open 
fitting may be detected by the REM reference microphone, interfering with the 
sound-field equalization. For this reason, the hearing instrument must be muted 
during sound-field equalization. 
1)  Position the client and probe tube and enter audiometric data. See 
10: Real-Ear 
Measurement Setup
17: Speechmap ®
18: Speechmap Fitting Procedures
2)  Highlight and <PICK> Instrument and select Open from the list. 
3)  Highlight and <PICK> a REAR test. If this is the first Open test in a session, you 
will be prompted to mute or turn off the hearing instrument and select the 
Equalize button on the setup poster. Otherwise, you are reminded to equalize 
whenever the client or nearby objects are moved. After equalization, unmute or 
turn the instrument on and follow the fitting procedures in 
18.5: Fitting to targets 
for soft speech
 to 
18.7: Fitting to targets for mid-level speech.