20  Troubleshooting 
20.1  Section overview 
FastFacts 20.1: Troubleshooting 
20.2  Self test failures 
FastFacts 20.2: Self test failures 
Any failure during power-on self test (P.O.S.T.) is indicated on-screen and further 
use of the instrument is denied. Try the following steps to resolve the failure: 
1)  To access Self test, press <Setup>, then highlight and <PICK> [Self test]. 
2)  If you received a Max out/in or a Routing failure, remove any hearing aids from 
the test chamber, and rerun self test. 
3)  Unplug all microphones (REM & HIT). Rerun self test. If self test passes, 
reconnect the microphones, one by one, running self test each time until a 
failure occurs. If the failure returns, display and print the test results and 
contact Audioscan for help. 
This section covers common problems encountered when using the RM500SL. If 
you require further assistance, note your model and serial number (located on the 
recessed panel) and contact your local distributor, or Audioscan at 519-268-3313 
800-265-2093 (USA only), 519-268-3256 (fax) or service@audioscan.com 
PLEASE DO NOT SHIP YOUR UNIT BACK TO THE FACTORY. In most cases 
we can courier you a part that you can quickly and easily replace yourself. 
This section covers the following topics: 
20.2: Self test failures 
20.8: Invalid HIT calibration 
20.3: HIT high distortion or noise 
20.9: No REM ref. mic. detected 
20.4: HIT curves inconsistent 
20.10: Invalid REM calibration 
20.5: HIT curves differ from specifications 
20.11: REM speaker overdriven 
20.6: HIT speaker overdriven 
20.12: Barcode scanner malfunction 
20.7: No HIT reference mic. detected