35 Default Settings Appendix
Alarm Default Settings
370
Alarm Default Settings
ECG, Arrhythmia, and ST Default Settings
Alarm Settings
Factory Default H10 / H20 / H40
H30 (deviations from H10/20/40)
Alarm Volume
5
Alarms Off
3 min.
Pause Al. 5min
Enabled
Pause Al. 10min
Enabled
Auto Alarms Off
Off
Alarm Off Reminder
Off
Visual Latching
Red & Yell 
Red Only
Audible Latching
Red & Yell 
Off
Alarm Reminder
On
Reminder Time
3 min
Alarm Sounds
Traditional 
Red Alarm Interval
10 sec
Yel. Al. Interval
20 sec
Alarm Low
4
2
Red Alarm Volume
AlarmVol +2
Yell. Alarm Volume
AlarmVol +0
Inop Volume
AlarmVol +0
Auto Increase Vol.
2 Steps
Increase Vol Delay
20 sec
Keep Blinking
No
Relay 1 Sensitiv.
R & Y & I
Relay 2 Sensitiv.
Red & Yell
Relay 3 Sensitiv.
Red
Inop Relay Latency
5 sec
Yel. Relay Latency
2 sec
Alarm Text
Standard
No Centr Mon Min Vol
4
ECG Settings
Factory defaults H10 / H20 / H40
H30 (deviations from H10/20/40)
Adult
 Pedi
 Neo
Adult
 Pedi
 Neo
High Limit
120 bpm
160bpm
200 bpm
Low Limit
50 bpm
75 bpm
100 bpm
Alarms
On
Alarm Source
HR
Auto
ECG
On
QRS Volume
1